In the event of failure, the following error codes will be indicated on the display:
Priority acc. to IEC 60601-1-8:2006
I
low priority
II
medium priority
ALARM- AND ERROR CONCEPT
Error category
Priority
Error
number
Indication in the
display
Description
10
II
101
Pressure too high
Internal pressure too high( >
90 mbar).
10
II
102
Sensor defective
O2 flow sensor defective
10
II
103
Sensor defective
Air flow sensor defective
10
II
104
No flow
Flow rate is zero
10
I
151
Flowrate not
reachable
Measured flow is lower then
the set flow ( about 2 l/min,
depending on the set flow)
10
I
153
Flowrate too high
Measured flow is higher that
the set flow ( about 2 l/min,
depending on the set flow)
10
I
154
Leakage detected
Leakage:
Humidification chamber not
available or leaking humidifica-
tion chamber.
10
I
155
Ambient pressure
off limits
Ambient pressure off limits
(limits see Chapter 7.3)
10
I
156
O
2
flow rate to high the added O
2
flow rate is to
high
10
I:
157
Sensor defective
Pressure sensor defective
20
I
251
Ambient
temperature
Ambient temperature out of
permitted range
(limits see Chapter 7.3)
20
I
252
Ambient
humidity!
Ambient humidity out of
permitted range
(limits see Chapter 7.3)
20
II
201
Air temperature
too high
Internal air Temperature to high
(> 43°C).
20
I
254
Sensor defective
Ambient temperature / humid-
ity sensor defective