Settings incorrect for Range Use Titan Interface software
to range the analog output
flow min and max points
Titan Interface Software Opens but
shows not connected to meter (no
spinning bars in bottom left of screen)
Two windows of the
Interface Software are open
and second window is
masking connected window
Either drag window out of
the way or close second
window using task bar
Damage to USB
Connection
Return to Titan Enterprises
for repair
Titan Interface Software is not
showing all Tabs (User Settings; Flow
Details; Advanced; Datalogging)
Software is placing the tab
window outside the monitor
area
Go to View on the top bar
menu and click Restore
Defaults. This restores the
Viewing defaults and re
docks all tabs to the main
window
Moisture in Display Window
Condensation or water in
the flowmeter
Dry unit before use
Extensive water
contamination will cause
multiple errors on the device
and will require repair at
Titan Enterprises
Leak from end of the body of meter
Poor seal on inlet/outlet
Correct seal
Housing damage on
polymer end caps leading
to internals leaking
If damage is seen, return
unit to Titan for repair and
recalibration