29
Titan Enterprises Ltd ©2021
Atrato_Instruction_Manual_0322
®
Once the Analog Mode selection has been
made you can enter the low and high flows
for the analog outputs as required. These are
numeric entries in the units and time base
selected in calibration settings.
4.4. Flow Details Tab
The Flow Details tab displays the performance of the meter in real time.
The main large display reflects the meter readings showing on the Atrato LCD Display, if fitted,
and will mimic that display. The buttons either side have the functions selected on the
‘Configuration’ screen and work in parallel with the buttons on the Atrato flowmeter
display unit
itself.
The flow displayed will be the Filtered Flow Rate as determined by the Viewing Filter Level in
the User Settings Tab.
The Results Details small window below the rate and total display shows various relevant
operational parameters whilst the unit is operating, e.g., relay status and frequency output if
these options are selected.