TE-5200
24
Operations Manual
Troubleshooting
*note: this is a general troubleshooting guide, not all problems may apply to
every sampler*
Problem
Solution
Brush Motor Won’t Turn On
-Check Motor brushes(Change every 500 hours)
-Check Motor(Should be replaced after 2 brush changes
about 1500 hours)
-Check power supply
-Ensure that all electrical connections are secure
-Make sure timer is on
-Make sure flow controller(if applicable) is adjusted
properly
-Check for loose or damaged wires
Brushless Motor Won’t Turn
On
-Ensure that all electrical connections are secure
-Make sure flow controller(if applicable) is adjusted
properly
-Check power supply
-Make sure timer is on
-Check for loose or damaged wires
Mechanical timer not working
-Make sure trippers are set properly
-Make sure that trippers are not pressed against switch at
start up, the timer need to rotate a few degrees before the
trippers hit the switch
-Check for loose or damages wires
-Check power supply
-Check electrical hook up diagram to ensure correct
installation
-Check Motor
Digital timer not working
-Check timer settings
-Make sure current date and time are correct
-Make sure power cords are properly connected
-Check fuse on main PC board (F3)
-Check Power Supply
-Check Motor
Summary of Contents for TE-5200
Page 2: ...TE 5200 2 Operations Manual TE 5200...
Page 31: ...TE 5200 31 Operations Manual Assembly Drawings TE 5005 Blower Motor...
Page 32: ...TE 5200 32 Operations Manual TE 5003 Filter Holder...
Page 33: ...TE 5200 33 Operations Manual Calibration Worksheet Sample...
Page 34: ...TE 5200 34 Operations Manual Calibration Certificate...