TE-5200
16
Operations Manual
∑
= sum of
Example Problems
The following example problems use data from the attached calibration
worksheet.
After all the sampling site information, calibrator information, and
meteorological information have been recorded on the worksheet, standard air
flows need to be determined from the orifice manometer readings taken during
the calibration using the following equation:
Qstd = 1/m[Sqrt((H
2
O)(Pa/760)(298/Ta))-b]
where:
Qstd = actual flow rate as indicated by the calibrator orifice, m
3
/min
H
2
O = orifice manometer reading during calibration, in. H
2
O
Ta = ambient temperature during calibration, K ( K = 273 + C
o
)
298 = standard temperature, a constant that never changes, K
Pa = ambient barometric pressure during calibration, mm Hg
760 = standard barometric pressure, a constant that never changes, mm Hg
m =
Qstandard slope of orifice
calibration relationship
b =
Qstandard intercept of orifice
calibration relationship.
Note that the ambient temperature is needed in degrees Kelvin to satisfy the
Qstd equation. Also, the barometric pressure needs to be reported in
millimeters of mercury and it has to be at the site pressure.
The following conversions may be needed:
2.
degrees Kelvin = [5/9 (degrees Fahrenheit - 32)] + 273
or degrees Kelvin = degrees C 273
3.
millimeters of mercury = 25.4 * (inches of Hg)
Inserting the numbers from the calibration worksheet run point number one
we get:
4.
Qstd = 1/1.47574[Sqrt((5.7)(757/760)(298/293)) - (-.00613)]
Summary of Contents for TE-5200
Page 2: ...TE 5200 2 Operations Manual TE 5200...
Page 31: ...TE 5200 31 Operations Manual Assembly Drawings TE 5005 Blower Motor...
Page 32: ...TE 5200 32 Operations Manual TE 5003 Filter Holder...
Page 33: ...TE 5200 33 Operations Manual Calibration Worksheet Sample...
Page 34: ...TE 5200 34 Operations Manual Calibration Certificate...