TE-5200
14
Operations Manual
standard
flows rather than actual flows and is taken from the Qstandard
section of the Orifice Calibration Certificate. The Qactual flows are used when
calibrating a PM-10 sampler.
The five orifice manometer readings taken during the calibration have been
recorded in the column on the data worksheet titled H
2
O (in). The five sampler
manometer readings taken during the calibration have been recorded under
the column titled FLOW (mano).
The orifice manometer readings need to be converted to the standard air flows
they represent using the following equation:
Qstd = 1/m[(Sqrt((H
2
O)(Pa/760)(298/Ta)))-b]
where:
Qstd = actual flow rate as indicated by the calibrator orifice, m
3
/min
H
2
O = orifice manometer reading during calibration, in. H
2
O
Ta = ambient temperature during calibration, K ( K = 273 +
o
C)
298 = standard temperature, a constant that never changes, K
Pa = ambient barometric pressure during calibration, mm Hg
760 = standard barometric pressure, a constant that never changes, mm Hg
m =
Qstandard slope of orifice
calibration relationship
b =
Qstandard intercept of orifice
calibration relationship.
Once these standard flow rates have been determined for each of the five run
points, they are recorded in the column titled Qstd, and are represented in
cubic meters per minute.
The sampler manometer readings taken during the calibration need to be
corrected to the current meteorological conditions using the following
equation:
FLOW (corrected) = [Sqrt((H
2
O)(Pa/760)(298/Ta))]
where:
flow (corrected) = sampler manometer readings corrected to current Ta and Pa
H
2
O = sampler manometer readings during calibration
Pa = ambient barometric pressure during calibration, mm Hg.
Summary of Contents for TE-5200
Page 2: ...TE 5200 2 Operations Manual TE 5200...
Page 31: ...TE 5200 31 Operations Manual Assembly Drawings TE 5005 Blower Motor...
Page 32: ...TE 5200 32 Operations Manual TE 5003 Filter Holder...
Page 33: ...TE 5200 33 Operations Manual Calibration Worksheet Sample...
Page 34: ...TE 5200 34 Operations Manual Calibration Certificate...