TE-5200
17
Operations Manual
5.
Qstd = .6776261[Sqrt((5.7)(.9960526)(1.0170648)) + .00613]
6.
Qstd = .6776261[Sqrt(5.7771295) + .00613]
7.
Qstd = .6776261[2.4 .00613]
8.
Qstd = .6776261[2.4096959]
9.
Qstd = 1.632
Throughout these example problems you may find that your answers vary
some from those arrived at here. This is probably due to different calculators
carrying numbers to different decimal points. The variations are usually slight
and should not be a point of concern.
With the Qstd determined, the corrected flow reading (FLOW (corrected)) for
this run point needs to be calculated using the following equation:
10.
FLOW (corrected) = [Sqrt((H
2
O)(Pa/760)(298/Ta))]
where:
FLOW (corrected) = sampler manometer readings corrected to standard
H
2
O = sampler manometer readings during calibration
Pa = ambient barometric pressure during calibration, mm Hg.
760 = standard barometric pressure, mm Hg
Ta = ambient temperature during calibration, K ( K = 273 +
o
C)
298 = standard temperature, K
Inserting the data from run point one on the calibration worksheet we get:
11.
FLOW (corrected) = [Sqrt(5.4)(757/760)(298/293)]
12.
FLOW (corrected) = Sqrt(5.47047)
13.
FLOW (corrected) = 2.3389035
This procedure should be completed for all five run points.
Summary of Contents for TE-5200
Page 2: ...TE 5200 2 Operations Manual TE 5200...
Page 31: ...TE 5200 31 Operations Manual Assembly Drawings TE 5005 Blower Motor...
Page 32: ...TE 5200 32 Operations Manual TE 5003 Filter Holder...
Page 33: ...TE 5200 33 Operations Manual Calibration Worksheet Sample...
Page 34: ...TE 5200 34 Operations Manual Calibration Certificate...