35
vsh-de-130326
3.3.2
Operation With Other Supply And Evaluation Units
The transducer can be operated with other customer related display units or
voltage supplies.
The electrical connection is to be made by means of suitable cables considering
EMI demands and according to the pin description shown below:
1
8
1
5
9
Socket
Pin 1:
Pin 2:
Pin 3:
Pin 4:
Pin 5:
Pin 6:
Pin 7:
Pin 8:
Pin 9:
Pin 10:
Pin 11:
Pin 12:
Pin 13:
Pin 14:
Pin 15:
Type SubD, 15pole, male
Identification: 39 k
Ω
Signal Output 0-10VDC
AGND
Voltage Supply 24VDC
Supply GND
Relay A, N.O.
Relay A, Common
Relay A, N.C.
Degas, extern
RS485 +
RS485 -
Shield
Relay B, N.O.
Relay B, Common
Relay B, N.C.
We recommend to have Shield (Pin 6) and supply common
(Pin5) grounded in the supply unit.
Incorrect connection or inadmissible supply voltage can damage
the transducer.