The unit requires no maintenance. External dirt and soiling can be removed by a
damp cloth.
Should a defect or damage occur on your VSH, please send the instrument to us
for repair and enclose a decontamination declaration.
The unit is not prepared for customer repair!
Defective sensor heads can be exchanged on-site (spare parts
B_VSH88 or B_VSH89).
Malfunction of the unit, which is caused by contamination or
wear and tear is not covered by warranty.
Error messages and malfunction
Problem
Possible Cause
Correction
high measurement error
contamination, ageing,
extreme temperature,
maladjustment
readjustment
0.5V < output signal < 1.3V /
"000000" via RS485
pressure under range
(pressure < 5x10
-10
mbar)
output signal < 0.5V /
"1" via RS485
Status LED continuously red
defective unit or sensor
send unit for repair
Status LED continuously
orange
hot cathode filament 1
defective
if appropriate exchange
sensor
"5" via RS485
Code unknown
check sent telegram code
"7" via RS485
logic error
command cannot be
executed at the moment
zero adjustment not possible measurement error exceeds
possible range of
readjustment
send unit for repair