34
vsh-de-130326
When mounting the transducer avoid forced twisting or violent
opening. This can damage the VSH!
Overpressure in the vacuum system > 1 bar
Accidental or unintended opening of clamp elements under
stress can lead to injuries due to parts flying around!
Unsecured hose connections can release, process media thus
can leak and possibly damage your health.
Overpressure in the vacuum system 1.5 to 4 bar
KF flange connections with elastomer sealings cannot withstand
such pressures. Process media thus can leak and possibly
damage your health.
3.3
Electrical Connection
3.3.1
Operation With Thyracont Display Unit
For operation of the transducer with a Thyracont display and control unit a
suitable measurement cable must be used (see accessories).
Do not connect or disconnect the transducer when the cable is
on circuit!
Connect the cables plug to the transducer and secure it with the screw. Connect
the other end of the cable to the display unit and secure the plug. Only now
connect your display unit to mains power or switch it on respectively.