4.10 Bake-Out – Detachment of Electronics
When a bake-out of the vacuum chamber is performed the temperature at the
sensor flange must not exceed 80°C.
Higher bake-out temperatures up to 180°C at the sen sor flange can be realized if
the electronics unit of the transducer is detached. For this purpose unscrew the
two knurled nuts as shown below and pull the electronics unit straightly and
without twisting off the flange.
The transducer must be disconnected from voltage supply before
detaching the electronics unit!
Make sure that no moisture or dust can intrude the electronics
unit!
When attaching the electronics unit pay attention to correct
orientation, do not twist while attaching!
This can damage the transducer!