KATflow 150
5 Commissioning
Defaults
Reload factory default settings, except for date
and time
Yes
No
Key Lock
Locks the keypad until password is entered (four
number keys followed by “ENTER”).
See also “Password” above.
Diagnostics
Temperature
Shows control unit temperature
Log Memory
Percentage of unused datalogger memory,
estimated time remaining
Datalogger
Interval
A value of zero turns the datalogger off, a non-
zero value turns the datalogger on and defines
the logging interval.
0 ... 999 s
Channel 1,
Channel 2
Selection
Select up to 10 items from list ↑↓
ENTER to select, 0 to remove
m/s, f/s, in/s, m3/h, m3/min, m3/s, l/h, l/min, l/s,
USgall/h, USgall/min, USgall/s, bbl/d, bbl/h,
bbl/min, g/s, t/h, kg/h, kg/min, m3, l, Usgall, bbl, g,
t, kg, W, kW, MW, J, kJ, MJ,
Sig dB (signal), noise dB, SNR,
C m/s (sound speed), CU (housing temperature)
Tin, Tout (inlet and outlet temperature)
TEMP (specified or measured fluid temperature),
SOS, DEN. KIN (derived sound speed, density,
kinematic viscosity)
Math (Calculated value – see below)
Low Memory
Warning output: The amount of memory remain-
ing at which the flowmeter begins to give an aud-
ible warning.
0 ... 100 %
Log Wrap
Saves "selected" items as a continuous stream
without headers (Note : this means files cannot
be processed by )
Yes/No
Log Download
Sends logger content to selected serial
communication port.
Log Erase
Erase datalogger
Yes/No
Serial Comms
Mode
Select from list ↑↓
None
Printer, Diagnostic, Log download,
Calibration Test (not normally used by user)
Baud
Select from list ↑↓
9600, 19200, 57600,115200
Parity
Select from list ↑↓
None
Even
(Default)
Odd
Type
Select from list where fitted
33