© 2016 Thorlabs
6 Communication with a PC
43
6.5.4 Service Request Enable Register (SRE)
The bits of the SRE are used to select which bits of the STB shall assert an SRQ.
Bit 0, 1, 2, 3, 4, 5 and 7 of the STB are combined by logical "AND" with the according 7 bits of the
SRE. These 7 results are combined by logical "OR", so that any "hit" leads to a logical 1 in bit 6 of
the STB and asserts an SRQ.
6.5.5 Reading the STB by Detecting SRQ
If an
is asserted, bit 6 of the STB is set to logical 1, so that the controller can detect by auto
serial polling, which device asserted the SRQ.
6.5.6 Reading the STB by *STB? Command
If the controller does not "listen" to SRQs at all, the service request can be detected by reading the
status byte with the command
"*STB?"
.
If bit 6 is logical 1, a service request was asserted.
6.5.7 Reading STB by Serial Poll
If the controller does not support auto serial poll, the service request can also be detected via
manual serial poll.
If bit 6 is logical 1, a service request was asserted.
6.5.8 Device Error Condition Register (DEC)
The bits of this register show the errors, that occur during operation (operation errors). The bits are
active-high.
If the error disappears, the bits are reset to LOW.
For TED8000 modules the bits 0, 4, 5, 6 and 8 are used
Bit 0 - Over temperature
TED8000 is overheated. Wait until the module has cooled down. Maintain proper air flow.
Bit 4 - Temperature out of window
The temperature of the device under control is out of specified window.
Bit 5 - Open circuit
The TEC circuit is open (interrupted connection).
Bit 6 - No or wrong Sensor
A temperature sensor could not be recognized, or the type of the recognized sensor does not
match the setting of the TED8000.
Bit 8 - Power supply error
Internal power supply error.
The DEC can be read but not set. Reading does not clear the DEC.
43
Summary of Contents for TED8000 Series
Page 1: ...PRO800 PRO8000 4 Series Modular Temperature Controllers TED8000 Operation Manual 2016 ...
Page 2: ...Version Date Item No 3 2 08 Sep 2016 M0009 510 227 Copyright 2016 Thorlabs ...
Page 42: ... 2016 Thorlabs 40 TED8000 Structure of the registers DEC DEE and EDE ...
Page 43: ... 2016 Thorlabs 6 Communication with a PC 41 Structure of the registers ESR ESE STB and SRE ...
Page 59: ......
Page 60: ...www thorlabs com ...