Portable Connector End Face Geometry Interferometer
Chapter 8: Image Tabs
Rev B, April 16, 2019
Page 38
Chapter 8
Image Tabs
8.1.
Live View Tab
Displays the live image of the connector being measured. When correctly focused, the live image shows the fringe
pattern that represents specific heights across the surface. The fiber should be completely inside the green circle.
If the fiber is not within the green circle, please conduct an Offset Calibration (see page 46.) or check that the mount
has been attached to the front panel of the CC6000 correctly.
Upon completion of a measurement, the live image will also display measurement details. These details include
the true size and location of the calibrated Apex Offset (indicated by the blue line from the fiber center terminating
in a red circle at true offset location), and the triangle symbol marking detected location of the fringe ring centers.
NOTE: Refer to Data Interpretation section in the Appendix for more detailed explanations of measured results
and calibrated measured results. (See page 58
.) If “Show Regions” option has been selected in the Results
Display Mode section of the Scan Options/Scan Settings Tab, the Measurement Regions will also be shown as
three concentric dashed circles on the live image.