![THORLABS Blueline Series Operation Manual Download Page 61](http://html.mh-extra.com/html/thorlabs/blueline-series/blueline-series_operation-manual_1110727061.webp)
The PRO8000 (-4) / PRO800 “ELCH” – macro functions
Pro8000 system / page
54
:ELCH:RUN 1
Reading:
:ELCH:RUN?
The measurement can also be started in triggered mode, that means you only start a
new measurement on the next step by the next query of a measured value:
Thus the user can affect measurement speed by the read frequency of measured
values, and also can stop the measurement interactively if limits of measured values
are reached.
Programming:
:ELCH:RUN 2
Reading:
:ELCH:RUN?
If the ELCH function is running in continuous mode at that moment, this command of
course triggers a short interruption in the measurement sequence.
The command:
:ELCH:TRIG?
Starts the measurement for the next step and reads the measured value
The order:
:ELCH:RUN 0
Will stop the measurement and discard the block function ELCH.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com