Hardware standard IEEE 488.1
Pro8000 system / page 25
[DCL]
Device Clear
The command [DCL] clears the output queue, the error queue and resets the
registers. It affects only the IEEE 488-bus electronic (see 3.6).
Normal operation of the modules is not interrupted.
NOTE
The command [DCL] will set back all IEEE 488 units connected to the bus.
[SDC]
Selected Device Clear
The command [SDC] resembles the command [DCL] but refers only to the IEEE 488
part of the selected unit (see 3.6).
Normal operation of the unit is not interrupted.
NOTE
In contrast to the command [DCL] the command [SDC] will only set back
the IEEE 488 of the device addressed.
[GET]
Group Execution Trigger
Due to interface specification “DT0” (refer to section 3.2.1, “IEEE 488.1 subsets of
the PRO8000 (-4) / PRO800” on page 21) the command
[GET]
is not implemented.