System reset
Pro8000 system / page
46
3.6 System reset
There are mainly 5 different ways to reset the PRO8000 (-4) / PRO800 or parts of it.
All of them behave different.
•
Power on reset
•
Reset by device clear
•
Reset by clear status
•
Reset by software
•
Reset by interface clear
3.6.1 Power on reset
The following happens inside a PRO8000 (-4) / PRO800 after turning on the power:
•
The power on bit in the Event Status Register (ESR) is set.
•
The output queue is deleted, bit 4 (MAV)of the status byte is set to 0.
•
The error queue is cleared, bit 2 (EAV) of the status byte is set to 0.
•
The execution unit and the command queue is cleared.
•
The command decoder (parser) is reset.
•
All event registers are cleared.
•
The device error summary register is 0.
•
All event registers are cleared.
•
System goes into 'ready' state, bit 0 (FIN) of the status byte is set.
•
All 'ELCH' macro functions are discarded.
•
All module outputs are switched off.
•
Set values are reset to the manually entered ones. Very new modules reset
these values to default values.