Error messages
Pro8000 system / page
34
3.4.1 Error queue empty
[
0,"No errors"
]
Error queue is empty.
3.4.2 100 ... 199 command error messages
[
100,"Unknown command"
]
Possible reason:
":HELLO WORLD". This string sent to the PRO8000 (-4) / 800
was not recognized as valid command.
":ILD:SET? 1.1". This command requires no numerical
parameter.
[
101,"Invalid character"
]
Possible reason:
Hex 09. This character sent to the PRO8000 (-4) / PRO800 does
not belong to the allowed set of characters.
[
102,"Invalid numeric parameter"
]
Possible reasons: ":ILD:SET 1.1." The second decimal point is not allowed.
":ILD:SET 12E+12E". The second "E" is not allowed.
[
103,"Invalid text parameter"
]
Possible reasons: ":SENSOR CG". This parameter is not valid for a temperature
sensor.
":MODE THH". This parameter is not valid for this command.
[
104,"Missing parameter"
]
Possible reason:
":ILD:SET". This command requires a numerical parameter.
[
105,"Invalid separator"
]
Possible reason:
":ILD:SE". This command was not complete.
[
107,"Empty slot"
]
Possible reason:
":SLOT 3" selected an empty slot.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com