The equipment must be set up for subject inspection in a temperature controlled
indoor environment. The location should be free from draughts, heater or air
conditioning ducts, and away from doorways. An environmental temperature
between 20°C and 24°C (68°F to 75°F) and Relative Humidity between 10% and 50%
should be maintained.
Instructions for arranging the camera, blackbody and person screening distance are
detailed in the Instruction Manual and should be observed. It is important that the
subject screening point is aligned with the location of the ThermaRef 35B1 blackbody
to ensure both are at the same distance from the camera focus point (1
–
3m, 3.2ft
–
9.5ft).
In the event that the equipment must be moved from one location to another, close
down the software on the computer first, then power off each item individually, dis-
connect all cables and return them safely to the heavy-duty transport case provided
with the system. Re-assemble in accordance with the directions stated in the
Instruction Manual.