05/2021, JK 

 

Backlight

  

When the instrument is operating in low-light conditions a photocell will detect this situation and turn the backlight 
on.

 

 
Temperature Logging 

It is necessary for the temperature to be stable before logging temperature. The display arrows will flash 3 times when 
the temperature reading has stabilized.    

 

If you wish to 

log the temperature

, Press the “down” button once until 

LOG

 is displayed. Press the “Select” button 

once to save a reading. An acknowledgment of a saved reading will occur with a display of “

LOG 1-10

”. This can be 

repeated for up to 10 saved readings.   
 
The logged temperatures can be accessed from the MEMORY menu. Logged temperatures will display with the LOG 
icon and a prefix (1,2,3,4,5,6,7,8,9,0) corresponding to the previously logged readings.  The up/down buttons can be 
used to advance through the readings. Select 

EXIT

 to return to the temperature display.  

 
Logged readings will be retained even after the unit is powered off. Logged readings can be cleared by going to the 
Memory section of the Menu and selecting 

CLR LOG

. New readings cannot be taken until the readings are cleared. 

 
USB Memory 

A micro-usb connection is available on the circuit board allowing access to calibration & logged data.  
Warning: Do not access the circuit board in a hazardous location.  
 

Lowest, Highest & Average readings 

The lowest, highest and average readings can be accessed through the MEMORY menu. The readings are indicated 
by the LOW, AVG or HIGH icons on the display. These readings are not related to the logged readings, but are 
determined from the temperature when the unit is powered on. These readings are deleted after the unit is powered 
off. 
 

Display Codes  

OPEN CKT

 - Indicates the sensor is operating above its temperature limit, the Probe Assembly is open circuited 

from a cut or broken section, or the cable is not properly inserted at the circuit board terminal. The most common cause 
is a damaged cable. 

  

SHORT CKT

 - Indicates the sensor is operating below its temperature limit or the Probe Assembly is short circuited 

due to a smashed or cut section. The most common cause is a damaged cable. 

  

NONE

 

– There are no logged readings saved in memory. 

 

2 WIRE

 

– A 2 wire probe connection has been detected.

 

3 WIRE

 

– A 3 wire probe connection has been detected. 

NO CAL

 

– The device does not have stored calibration data for the temperature probe. Perform an  

adjustment/calibration before use. 
 
 

 
 
 
 
 
 
 
 
 
 
 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for TP7-D

Page 1: ...Thermoprobe Models TP7D and TP9A TP7 D TP9 A USER MANUAL www com information itm com 1 800 561 8187...

Page 2: ...tified Body Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul Naviglio MI Italy Identification Number 2575 Auditing body QAN Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul N...

Page 3: ...rsed where one cell could charge another cell New batteries must not be mixed with old batteries Batteries must not be mixed with batteries of other models or manufacturers a Ensure the instrument is...

Page 4: ...up to 10 readings C F Select Celsius or Fahrenheit temperature display DECIMAL Select 0 1 or 0 01 display resolution MEMORY Use up down buttons to display Lowest reading Average reading Highest readin...

Page 5: ...USB Memory A micro usb connection is available on the circuit board allowing access to calibration logged data Warning Do not access the circuit board in a hazardous location Lowest Highest Average r...

Page 6: ...a through c to remove batteries b On the circuit board push the terminals clamps down and remove the wires noting the wire lead color code arrangement See Figure 1 White positive sensor wire Green cab...

Page 7: ...ence thermometer Note for limited calibration If entire range capability of instrument is not required the 2 point high adjustment can be made at a temperature relatively close to the common temperatu...

Page 8: ...return the unit to a factory calibration curve calibration adjustment will still be required o The RESETCAL reset to factory calibration feature should be used if the probe is replaced using a 2 wire...

Page 9: ...nd shall only be undertaken by ThermoProbe or one of its qualified service providers See also Authorized Repair section g To reduce the risk of fire or explosion this device must be bonded to the vess...

Page 10: ...must be present An intrinsically safe system assumes the fuel and oxygen is present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument...

Reviews: