background image

05/2021, JK 

 

3.

 

The device is now in adjustment mode.  The display will show 

ADD PT1

. The “up/down” buttons can be used 

to select a different point to adjust or 

CANCEL

. If the 

CANCEL

 option is chosen then the calibration 

procedure is exited and the prior calibration values are re-activated.    

ADD PT1

 = Lowest temperature point 

ADD PT2 

= the next higher temperature point   

ADD PT3 

= the next higher temperature point (only used in 3-point calibration or 4-point calibration mode)   

ADD PT4 

= the highest temperature point (only used in 4-point calibration mode)   

ADJ DONE 

= Save and exit calibration mode (all points adjusted & valid)  

CANCEL

 = Exit calibration mode without saving 

 
Pick which point to adjust and press the select button to begin. The current temperature will display and the 
CAL icon and the number of the point to adjust will now be blinking.  

 

NOTES:  

o

 

Calibration can be performed to hundredths of a degree.   

o

 

The up/down buttons can be used to increase or decrease the display reading.   

o

 

Holding 

the up/down buttons adjusts 

0.1 degrees 

increments.   

o

 

Momentary Presses 

of the up/down buttons for less than 0.5 seconds adjusts 

0.01 degrees 

for every 

press.   

o

 

The display arrows will flash 3 times when the temperature reading has stabilized. 

o

 

All points must be saved before selecting 

ADJ DONE

DATA ERR

 message will display if all 

points have not been saved or are not in increasing temperatures. (Ex: PT1 = 32F, PT2 = 120F, PT3 
= 250F) 

o

 

While in calibration mode the temperature will display based on the prior calibration curve. 

o

 

The 

RESETCAL

 feature can be used to return the unit to a factory calibration curve. 

(calibration/adjustment will still be required). 

o

 

The 

RESETCAL

 (reset to factory calibration) feature should be used if the probe is replaced using 

a 2-wire assembly 

instead

 of a 3-wire assembly or if prior calibration has caused the unit to read 

excessively out of tolerance. 

 

4.

 

 Once the temperature has stabilized, using a reference device check the actual temperature in the bath and use 
the “up/down” buttons to adjust the device to the actual temperature. Once the device temperature matches the 
actual temperature, 

press the select button to save the setting

. The display will show 

SAVE

 or 

EXIT

. If 

EXIT

 is selected it will return to the 

ADD PT1

 menu. Once 

SAVE

 is selected the display will advance to the 

next temperature point for adjustment showing 

ADD PT…

The up/down buttons can be used to select a 

different point to adjust or 

CANCEL

. If the 

CANCEL

 option is chosen then the calibration procedure is exited 

and the prior calibration values are re-activated.   

o

 

NOTE: Once 

SAVE

 is selected the temperature display will return to the previous reading until the 

calibration is completed. 

o

 

NOTE: If 

SAVE

 is selected before the temperature has stabilized, the display will show 

NOT 

STABLE

. Wait for the temperature to stabilize before saving. 

 

5.

 

Move the probe to the next bath and repeat step 4. After you save the highest temperature point the display 
will flash 

READY

 and the new calibration settings will be in effect. The buttons will now resume their 

normal operating functions. The calibration settings are saved to flash memory when the device is turned off. 
The unit will not turn off automatically. Manually turning the unit off saves the calibration data.   

 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for TP7-D

Page 1: ...Thermoprobe Models TP7D and TP9A TP7 D TP9 A USER MANUAL www com information itm com 1 800 561 8187...

Page 2: ...tified Body Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul Naviglio MI Italy Identification Number 2575 Auditing body QAN Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul N...

Page 3: ...rsed where one cell could charge another cell New batteries must not be mixed with old batteries Batteries must not be mixed with batteries of other models or manufacturers a Ensure the instrument is...

Page 4: ...up to 10 readings C F Select Celsius or Fahrenheit temperature display DECIMAL Select 0 1 or 0 01 display resolution MEMORY Use up down buttons to display Lowest reading Average reading Highest readin...

Page 5: ...USB Memory A micro usb connection is available on the circuit board allowing access to calibration logged data Warning Do not access the circuit board in a hazardous location Lowest Highest Average r...

Page 6: ...a through c to remove batteries b On the circuit board push the terminals clamps down and remove the wires noting the wire lead color code arrangement See Figure 1 White positive sensor wire Green cab...

Page 7: ...ence thermometer Note for limited calibration If entire range capability of instrument is not required the 2 point high adjustment can be made at a temperature relatively close to the common temperatu...

Page 8: ...return the unit to a factory calibration curve calibration adjustment will still be required o The RESETCAL reset to factory calibration feature should be used if the probe is replaced using a 2 wire...

Page 9: ...nd shall only be undertaken by ThermoProbe or one of its qualified service providers See also Authorized Repair section g To reduce the risk of fire or explosion this device must be bonded to the vess...

Page 10: ...must be present An intrinsically safe system assumes the fuel and oxygen is present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument...

Reviews: