 USER INSTRUCTIONS - TP7-D & TP9-A 

     

 

INTRODUCTION  

 

This manual describes the basic function, use and safety instructions for a model TP7-D and TP9-A portable digital 
thermometer instrument. 

 

  

REPLACING BATTERY  

When the battery voltage is low, the low battery icon will show on the display.  
 
When the battery voltage is 

very

 low, the backlight will no longer operate and the low battery icon will ‘blink’ on the 

display.  
 
Replace batteries as soon as possible in a safe location after the low battery is noticed. This will ensure backlight 
operation, and avoid possible malfunctioning. Do not attempt to calibrate the instrument if the low battery indicator is 
displayed.  

 

WARNING:  

 

 

Batteries must be changed in Non-hazardous area.   

 

Batteries must be of correct approved type.   

 

Batteries must be installed with correct polarity making sure the (+) end of the battery is aligned with (+) 
symbol embossed in the battery case.  

 

Batteries must not be installed with polarity reversed where one cell could charge another cell.   

 

New batteries must not be mixed with old batteries. Batteries must not be mixed with batteries of other 
models or manufacturers.  
 

a) Ensure the instrument is in a non-hazardous area & powered off. 
b) Use a #2 Phillips drive to remove the 3 screws holding the front cover on the TP7-D or the 2 screws holding the 
front cover on the TP9-A.  
c) Use a #1 Phillips drive to remove the single screw from the battery cover. Remove the battery cover, push one 
battery towards the spring contact and lift battery up from the holder, and then remove the remaining battery.  
d) Install each new battery making sure the (+) end of the battery is aligned with (+) symbol embossed in the battery 
case.  
e)

 

Replace the retaining device and reinstall the cover.  

  

CERTIFIED

 Batteries for the TP9-A and TP7-D are as follows:  

 

Manufacturer    

Type  

 

 

Part Number 

 

Duracell  

 

AA (LR6) Alkaline  

MN1500 

Panasonic  

 

AA (LR6) Alkaline 

LR6XWA 

GP (Gold Peak)    

AA (LR6) Alkaline  

GP15A 

 
 

AUTHORIZED REPAIR 

 

It is recommended that service beyond the scope of this manual be performed by ThermoProbe, Inc. or one of its 
authorized distributors.  

 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for TP7-D

Page 1: ...Thermoprobe Models TP7D and TP9A TP7 D TP9 A USER MANUAL www com information itm com 1 800 561 8187...

Page 2: ...tified Body Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul Naviglio MI Italy Identification Number 2575 Auditing body QAN Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul N...

Page 3: ...rsed where one cell could charge another cell New batteries must not be mixed with old batteries Batteries must not be mixed with batteries of other models or manufacturers a Ensure the instrument is...

Page 4: ...up to 10 readings C F Select Celsius or Fahrenheit temperature display DECIMAL Select 0 1 or 0 01 display resolution MEMORY Use up down buttons to display Lowest reading Average reading Highest readin...

Page 5: ...USB Memory A micro usb connection is available on the circuit board allowing access to calibration logged data Warning Do not access the circuit board in a hazardous location Lowest Highest Average r...

Page 6: ...a through c to remove batteries b On the circuit board push the terminals clamps down and remove the wires noting the wire lead color code arrangement See Figure 1 White positive sensor wire Green cab...

Page 7: ...ence thermometer Note for limited calibration If entire range capability of instrument is not required the 2 point high adjustment can be made at a temperature relatively close to the common temperatu...

Page 8: ...return the unit to a factory calibration curve calibration adjustment will still be required o The RESETCAL reset to factory calibration feature should be used if the probe is replaced using a 2 wire...

Page 9: ...nd shall only be undertaken by ThermoProbe or one of its qualified service providers See also Authorized Repair section g To reduce the risk of fire or explosion this device must be bonded to the vess...

Page 10: ...must be present An intrinsically safe system assumes the fuel and oxygen is present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument...

Reviews: