background image

Models TGO-80W, 150W, 240W             Glassware Drying Oven Manual 

26-1-16 

2:

  

Power switch:   

 

 

 

 
This illuminated switch is used to turn on the temperature controls and the 

circulating fan. 

“0” 

is power off and 

“1” 

is power on. 

 

 
3:

 

Temperature control: 

 

 

PV (process variable): 

This digital display is the actual temperature in the oven. 

 

SP (set point): 

 

This digital display is the set temperature in the oven. 

 

 

This button is used to enter changes to parameters and to 
move between functions. 

 

 

 

These arrows are used to move between, and to alter the 
numbers on the screen. 

 
4: 

Timer Switch:

 

 

 

 

This switch turns the duration timer on in position 

“1” 

or allows heating to be on 

continuously in position 

“0”

. The indicating light above the switch will be on when the set 

duration time is complete. 
 

Summary of Contents for TGO-150W

Page 1: ...Scientific Equipment Pty Ltd ABN 80 000 859 129 10 12 Ross Place Wetherill Park NSW 2164 P O Box 1851 Wetherill Park DC NSW 1851 Phone 02 9604 3911 International 61 2 9604 3911 Fax 02 9725 1706 Intern...

Page 2: ...the packing material and check for damage Notify the detail of any damage to your supplier or to Thermoline Scientific without delay Retain the packing materials until the equipment has been thorough...

Page 3: ...for the oven The TGO 80W and the TGO 150W are suitable for bench mounting but the larger TGO 240W is meant to be free standing Model TGO 80W TGO 150W TGO 240W Internal Dimensions 500 x 400 x 400 WDH...

Page 4: ...he actual temperature in the oven SP set point This digital display is the set temperature in the oven This button is used to enter changes to parameters and to move between functions These arrows are...

Page 5: ...ure will go through a short warm up before displaying the set temperature SP on the lower display and the oven temperature on the top display PV Use the arrow to highlight the digit on the set tempera...

Page 6: ...ached the heating will turn off and the indicating light above the timer switch will be on Note to reset the timer you have to turn the main power switch off Press the button again to return to the or...

Page 7: ...t accidental overheating Please be aware that the thermostat is a simple device that is less accurate than the digital controls It should not be set too close to the operating temperature in case it i...

Page 8: ...the oven by a certified instrument For this reason the digital temperature control has a parameter that can be used to correct the temperature displayed called the BS In simple terms this BS adds or s...

Page 9: ...ke the LOCK parameter 1 again before exiting the parameter configuration level by holding the button for a few seconds The output of the temperature control will now adjust the oven to the correct cal...

Page 10: ...major failure Notwithstanding the preceding clause and to the extent permissible by law the liability of Thermoline is limited in relation to the warranted product and at the option of Thermoline to r...