In the Miscellaneous box, the use of a foot pedal switch (as alternative for the stylus on the
touchscreen), and the possibility to switch off the humidifier during manual application and
plunge freezing can be chosen. The semi-automatic grid transfer (i.e. the automatic
movement of the grid from the liquid ethane/propane towards the grid box in the liquid
nitrogen atmosphere) is default activated but can be de-activated by checking the skip box. In
case the auto raise ethane lift option is checked, the coolant container will be lifted towards
the bottom of the climate chamber simultaneously with the uplift of the tweezer (you can also
do this in separate actions).
In addition, all essential freezing parame
ters can be saved (‘Save’) and loaded (‘Load’)
according to the type of sample or experiment that is performed.
Summary of Parameters that can be set
Parameter
Range
Increment
Blot-time (s)
0 - 99999
± 0.1
± 0.5
Blot process
0 - 16
1
1
Blot force
-25 - +25
± 1
± 1
Drain time (s)
0 - 99999
± 0.1
± 0.5
Wait time (s)
0 - 99999
± 0.1
± 0.5
Skip application process
on / off
Use foot pedal
on / off
Hum off during process
on / off ransfer
Skip Grid Transfer
on / off
Auto raise Ethane lift
on / off