Quantifoil or lacey carbon film is
damaged during processing
Centering of forceps between the blot pads is incorrect. This is likely
the result of either one of the two causes: 1 the tip of the forceps is
misaligned with respect to the rod catching brass part (accidental
dropping or undue force?) or alternatively 2 the blotting mechanism
is misaligned with respect to the vertical movements of the forceps.
Check mechanical centering of the tip of the forceps and the brass
part with the alignment tool (alignment tool is not standard delivered
with each system). If this is alignment is correct and problems
remain the mechanical alignment of the blotting mechanism is the
next probably cause of specimen damage. Consult the service
instructions.