SPECTRONIC 200E USER GUIDE
SPECTRONIC 200 Modern Interface
Thermo Scientific
SPECTRONIC 200E User Guide
13
7. Select
GO
and press
to restart the measurement.
Quantitative Analysis
The Quantitative Analysis (Quant) application supports both linear calibration curves with
up to four standards and manual entry of a factor for determination of the concentration.
Figure 15.
Use the arrow keys on the navigation pad to select the number of standards, Measurement
Mode (ABS or %T) and the display unit. Set the Measurement wavelength using
or
the
λ
knob.
The USB memory option allows you to select:
Save
– to automatically enter the method save sequence
No
– to go directly to sample measurement
Load
– to automatically enter the method load sequence
Once the method parameters are complete, select
GO
and press
.
Note
If you change the measurement wavelength, it may be necessary to re-zero the
instrument by placing a blank solution in the sample stage and pressing
until the
screen displays the message
Performing Auto Zero
.
0.00
Application
No. of STDs(F,1-4)
Measurement Mode
GO
Measurment
?
Unit
USB memory
Quant
ABS
1
340
nm
ppm
Save