RC-5C PLUS
Operation
4-11
Entry Errors
Any of the following conditions will cause the ENTRY ERROR
advisory message to light:
•
The MAX temperature switch is set at or below the value set on
the RUN temperature switch. In this case, the MAX temperature
that the system will use is the RUN temperature plus seven
degrees. Change the MAX temperature setting for a value at
least 5
°
C higher than the RUN temperature setting.
•
The RPM switches are set for a speed below 50 rpm. The system
will default to a 50 rpm setting. Change the speed setting.
•
The RPM switches are set at or above 105% of the maximum
speed of the identified rotor. Check the ROTOR CODE setting to
make sure you have selected the proper code number and check
the speed setting to make sure it is within range for the rotor.
•
The TIMED/HOLD switch is set at TIMED, the OFF/
∫ω
2
dt
switch
is set at OFF, and the TIME (HRS and MIN) switches are set at
zero. The run will not start.
•
The TIMED/HOLD switch is set at TIMED, the OFF/
∫ω
2
dt
switch
is set at
∫ω
2
dt
, and the COEF switch is set at zero. The run will
not start.
•
The ROTOR CODE switches are set at a number that is not used
(see Table 4-1 for numbers that are used); the control values will
default to those for a ROTOR CODE setting of “00”.
4-11/4-12