The Main Screen
7 of 73
1.2.1
Time
The run time can be changed via the Time button.
Here, you can enter a run time in hours and minutes. (You can also activate an input for seconds by making the
appropriate change in the settings.)
There are three different modes for the run time.
1-
Time
A time can be entered. When this time expires, centrifugation stops. You can specify whether the set time
begins when the centrifuge is first started
or once the acceleration phase is completed
.
2-
HOLD Mode
In HOLD mode, no time is entered. Instead, centrifugation continues until it is manually stopped.