The Main Screen
15 of 73
1.3
Centrifuge Control Elements
The bottom area of the main screen can be used to control the centrifuge.
Various buttons can be displayed in the centrifuge control area:
The green Start button starts the centrifuge using the current set values,
provided that the door is closed and a rotor is inserted.
If the door is open, this is automatically closed when the Start button is
pressed.
If the Timed Start button was activated when the program was selected,
this button replaces the green Start button (see Timed Start section).
The red Stop button halts the current centrifugation; the progress indicator
moves to the deceleration phase.
If a timed start has been planned and the start time has not yet arrived,
pressing the Stop button cancels the timed start.
Opens the door of the centrifuge.
Closes the door of the centrifuge.
Cannot operate door. This status occurs if the centrifuge has been switched on
while the door is open. In this situation, the door must be closed manually,
after which automated control becomes available again.
Configuration
Centrifuge control elements
User login
Context-sensitive help
Summary of Contents for Sorvall BIOS 16
Page 43: ...Menu 43 of 73 ...