Smart2Pure Pro Ultrapure water system
System Control | 33
The different menu points in the OEM-Menu are listed in the
below.
Table 3. OEM Menu Options
Menu
Description
Settable
Range
Factory
Setting
Actions
OEM MENU
OEM Menu Prompt
N/A
N/A
Press ENTER key to enter into the
OEM Menu
Max Temp
Alarm set point for maximum
allowable temperature.
1 - 50°C
40°C
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Recirculation
Interval
Time setting for how long the
Smart2Pure Pro will
recirculate water in the
Dispense Mode
10, 20, 30
minutes
10 minutes
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Rinse Interval
When the unit switches from
standby mode into dispense
mode (or tank filling) a short
rinse of the RO cartridge
occurs
0.1 - 10
seconds
2 seconds
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Date & Time
The date and time can be set
in this screen.
1-12 month,
1-31 days,
0-24 hours,
0-60minutes
N/A
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Print Interval
If printer option is used, this
menu is where to set the
frequency of the data
printout
5 minutes - 12
hours
5 minutes
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Language
Setting for different
languages
English,
French,
German
English
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
μS/cm -
MΩ*cm
Allows for water purity to be
read in conductivity (μS/cm)
or resistivity (MΩ*cm)
μS/cm -
MΩ*cm
MΩ*cm
Use Up/Down Arrow keys to adjust
setting and press ENTER to confirm
Change
Temp. comp.
Allow purity on the display to
be temperature
compensated to 25°C
ON - OFF
off
Use Up/Down Arrow keys to adjust
setting and press ENTER to confirm
Change
OFF/ON/
Empty
Tank level settings (OFF/ON)
to adjust when filling of the
tank stops (OFF) and when it
starts again (ON). Also, sets
the level at which the
recirculation pump can turn
on (Empty)
Off: 40 - 100%
On: 10-70%
Empty: 10-
25%
Off: 100%
On: 70%
Empty: 25%
Use Up/Down/Right Arrow keys to
adjust setting and press ENTER to
confirm Change
Leak
Detection
Turns on leak detection and
threshold. If ON, if leak is
detected all pumps and feed-
water solenoid valves are
de-energized.
ON - OFF
On
Use Up/Down Arrow keys to adjust
setting and press ENTER to confirm
Change