32 | System Control
Smart2Pure Pro Ultrapure water system
System Control
General Information
When the ON/OFF key is pressed, the system starts in the
Standby Mode, recirculating pump is off, the booster pump
comes on based on the tank level and level set points.
User Menu
User menu parameters monitor the operation of the system.
Use Menu button to access and scroll through User Menu
parameters.
below describes parameter, settable
range if adjustable, factory default setting and appropriate
action when required.
OEM Menu
Basic operation settings and limiting values can be changed
in OEM-menu. To make changes in the OEM-menu, the
system control must previously be unlocked.
1. Press Menu button until CODE is displayed.
2. Using Up/Down and over keys enter code 0450, Press
ENTER.
3. Simultaneously pressing the Up/Down keys. The prompt
“OEM-menu Press enter!” appears.
4. Use Menu button to scroll through parameters.
Table 2. User menu options
Menu
Description
Settable
Range
Factory
Setting
Actions
Pure Water Limit
Alarm Setpoint to alert
user to replace cartridge
1 - 10 MΩ*cm
2 MΩ*cm
Change DI Cartridge
RO Product
RO permeate
conductivity value (Purity
is only displayed in
conductivity μS/cm)
N/A
N/A
Monitor for changes in RO
permeate purity. If purity drops
below 70% of feed-water
conductivity, replace RO
cartridge.
UV Time Left
How many hours until the
UV bulb needs to be
replaced
3000 Hours
Change UV when UV Duration
alarm appears and reset timer
in Service Menu
RO time left
How many days until the
RO needs to be replaced
360 Days
Change RO cartridge at 0 and
reset timer in Service Menu
UF time left
How many days until the
UF needs to be replaced
720 Days
Change UF at 0 and reset timer
in Service Menu
DI time left
How many days until the
DI needs to be replaced
360 Days
Change DI at 0 and reset timer
in Service Menu
Error History
Alerts listed by date or if
menus were unlocked
N/A
Press ENTER to view error
history, and use the Up/Down
keys to scroll through the errors
Code
Unlocking code to allow
access to change values
in Service or OEM Menus
N/A
Input code 0450 and press