![Thermo Scientific Smart2Pure Pro Installation And Operation Manual Download Page 25](http://html.mh-extra.com/html/thermo-scientific/smart2pure-pro/smart2pure-pro_installation-and-operation-manual_1105419025.webp)
20 | Installation
Smart2Pure Pro Ultrapure water system
11. Quick connectors for RO cartridge
12. Quick connectors for ultrapure cartridge
Note:
The external terminals of the equipment are only
required to be connected with external circuits with the
voltages below the limits of 6.3.1 of 61010-1 or 30 V r.m.s.
and 42.4 V peak or 60 VDC and are electrically separated
from hazardous live voltages by reinforced or double
insulation.
Connections of the
Smart2Pure Pro Pure
Water Storage Tank
Figure 16. Water storage tank connection
13. Tank flow connector for outer diameter 6 mm tube
14. Level & Pump control cable with connector
15. Tank back flow connector for outer diameter 6 mm tube
16. Connector for required sterile overflow, outer diameter
8 mm tube which connects to the atmospherically vented
drain
17. Plug to be removed for required CO
2
Ab0.2
micron Vent Filter
18. ¾” NPT outlet connector for pressurized flow to remote
hand gun or equipment feed
Water Connections of
Smart2Pure Pro System
Make sure that you have installed upstream, the Smart2Pure
Pro system and the required pretreatment accessory
50157886 (triple cartridge pretreatment system).
Refer to the section
installation tips. To ensure a leak free connection, push the
tubing into the fitting 12 - 13 mm (½”) until the tubing locks in
place.
Note:
Use tube cutter to ensure ends of tubing are flat and
free of burrs (prickles).
Refer to
for assistance in making the connections
on the Smart2Pure Pro. If installing alternative pretreatment,
refer to the section
Alternative Feed Water Connections
Install system tubing in the following order:
Pretreatment connections:
Connect ¾” to ¼” feed-water connector from the assembly
kit, to the feed-water source.
Figure 17. Feed-water fitting
1. Connection to feed water supply requires Barnstead
Pretreatment system, 50157886 (refer to
Pretreatment
System User manual
for installation instruction),
Connect ¼” tube to feed-water connector and to
pretreatment “Feed-water Inlet”
2. Connect ¼” tube to pretreatment “Water Outlet”
Note:
Maximum distance between pretreatment and
Smart2Pure Pro should be 3 meters.
3. Connect ¼” tubing from pretreatment “Water Outlet” to
the feed-water connector on the Smart2Pure Pro
Connect ¼” tubing from assembly kit to the concentrate
connector (for drain)
Smart2Pure Pro and Tank Connections:
•
Connect 6 mm tubing from assembly kit to the Tank Flow
connector on Smart2Pure Pro (to be connected to tank)
13
14
15
18
16
17
Dirt Screen
¾” Union Nut
Outer Diameter
¼” Tube
Fitting