Chapter V
Electrode Care and Maintenance
For .best .results, .always .keep .the .pH/ORP .electrode .bulb .wet . .Store .the .pH/ORP .glass .
bulb .with .pH .electrode .storage .solution . .Other .pH .buffers .are .also .suitable . .NEVER .
use .deionized .water .for .storage . .Wash .the .probe .thoroughly .with .distilled .water .after .
each .use . .Because .your .pH .electrode .is .susceptible .to .contamination .or .dirt, .clean .it .
every .1 .to .2 .months .depending .on .the .extent .and .condition .of .use .
Clean .the .pH/ORP .electrode .using .a .mild .detergent . .Wipe .the .probe .with .a .soft .tissue .
paper . .Avoid .touching .the .glass .membrane .with .your .fingers . .Wash .thoroughly .in .tap .
water .and .then .in .distilled .water . .Recalibrate .the .meter .after .cleaning .the .electrode .