Thermo Scientific Russell RL060P pH Meter User Guide

Without Temperature probe (no ATC)

If .no .temperature .probe .is .used, .the .meter .compensates .for .pH .response .based .on .a .
new .calibrated .temperature .value .manually .set .by .you .or .at .25 .0 .°C .(factory .default) . .

1 . . Press .

 .to .enter .into .temperature .mode .until .“°C” .shows .in .LCD . .

2 . . Compare .the .displayed .value .to .NIST .certified .thermometer .or .a .thermometer .

known .to .be .accurate .(dipped .into .a .constant .temperature .bath) .

3 . . Press .

 .to .enter .temperature .calibration .mode . .The .LCD .shows .“

CA

” .

momentarily .and .displayed .reading .flashes . .Note .that .this .displayed .value .should .
either .be .25 .0 .°C .or .the .last .set .temperature .value . .

4 . . Press .

 .(for .pH .5/6) .until .the .displays .shows .the .desired .temperature . .You .can .

set .any .value .from .0 .to .100 .°C .

5 . . To .cancel .or .abort .this .operation, .press .

 . .Note .no .new .value .will .be .stored .

into .the .meter‘s .non-volatile .memory . .To .confirm .calibration, .press .

 . .The .LCD .

displays .“

CO

” .momentarily, .and .the .meter .reverts .to .measurement .mode .

Summary of Contents for Russell RL060P

Page 1: ...Russell RL060P User Guide...

Page 2: ...riode Quatrode QuiKcheK rf link ROSS ROSS Resolution SAOB SMART AVERAGING Smart CheK SMART STABILITY Stacked Stat Face The Enhanced Lab ThermaSense Triode TRIUMpH Unbreakable pH Universal Access are t...

Page 3: ...alibration 6 Chapter IV Measurement 14 Chapter V Electrode Care and Maintenance 15 Chapter VI Troubleshooting 16 Chapter VII Specification 17 Chapter VIII Ordering Information 18 Chapter IX Assistance...

Page 4: ......

Page 5: ...rge custom LCD Liquid Crystal Display for clear and easy reading The RL060P pH meter measures pH and temperature C Included with your meter are 4 alkaline AAA batteries sensor user guide and a warrant...

Page 6: ...m Selects measurement mode for mV pH and temperature Allows calibration for pH mV or temperature and will also abort the calibration without confirming any set value hold Freezes the measured reading...

Page 7: ...cted 2 To insert the meter into the rubber armor slide in from the top of meter before pushing the bottom edges down and set it into position Lift up the stand at the back of the meter for bench top a...

Page 8: ...ocks Do not force when connecting To remove simply rotate the connector in anti clockwise direction until it unlocks and slide the connector off the socket Insert the mini phono jack of the temperatur...

Page 9: ...rement The corresponding annunciator displays in the LCD In temperature mode the measured reading can be 25 0 C factory default The last calibrated temperature value will appear if there is no tempera...

Page 10: ...buffer recognition that identifies the correct pH buffer values during calibration If non standard pH buffers other than the above standards are used or the electrode has worn out the LCD will flash E...

Page 11: ...mode Press to toggle USA NIST or Pb standards 3 Press to confirm your choice of buffer standard to be used The display then reverts to the pH measurement mode Resetting the User Calibrated Values If y...

Page 12: ...ode condition 3 Press to enter pH calibration mode A CA displays momentarily and the display shows the current uncalibrated reading flashing while in the calibration mode 4 To abort or cancel calibrat...

Page 13: ...ode condition 3 Press to enter pH calibration mode A CA displays momentarily and the display shows the current uncalibrated reading flashing while in the calibration mode 4 To abort or cancel calibrat...

Page 14: ...our electrode condition 3 Press to enter pH calibration mode A CA displays momentarily and the display shows the current uncalibrated reading flashing while in the calibration mode 4 To abort or cance...

Page 15: ...tabilizes 3 Press to enter mV calibration The LCD shows CA momentarily followed by displayed reading flashes 4 To abort calibration press Meter reverts to measurement mode 5 To proceed with the calibr...

Page 16: ...pears in the LCD 2 Compare the displayed value to a NIST certified thermometer or other thermometer known to be accurate For best accuracy place both the probe and thermometer in a constant temperatur...

Page 17: ...ometer known to be accurate dipped into a constant temperature bath 3 Press to enter temperature calibration mode The LCD shows CA momentarily and displayed reading flashes Note that this displayed va...

Page 18: ...l gently and wait for the reading to stabilize Note the reading Freeze the display if necessary for details refer to the Releasing a Held Reading section 4 Rinse the probes with tap water or rinse wat...

Page 19: ...d water for storage Wash the probe thoroughly with distilled water after each use Because your pH electrode is susceptible to contamination or dirt clean it every 1 to 2 months depending on the extent...

Page 20: ...ones Unstable reading a Electrode not deep enough in sample b Dirty electrode c Broken electrode a Place electrode deeper in sample b Clean electrode and recalibrate c Replace electrode Er1 display B...

Page 21: ...rature Range 0 0 to 100 0 C Temperature Resolution 0 1 C Temperature Accuracy 0 5 C Temperature Compensation Automatic Manual 0 to 100 C Millivolt Range 1000 to 1000 mV Millivolt Resolution 1 mV Milli...

Page 22: ...with 3 in 1 pH ATC electrode 9147BN without rubber armor 9142BN Glass body refillable combination pH electrode single junction 12 mm x 120 mm 1 meter cable length 9146BN Epoxy body gel combination pH...

Page 23: ...les 700702 pH 6 97 low ionic strength buffer 4 x 475 mL bottles 910107 pH 7 00 buffer 475 mL bottle 1 pint 910760 pH 7 00 buffer 5 x 60 mL bottles 910710 pH 7 00 buffer pouches 10 x 15 mL pouches 9107...

Page 24: ...contact Technical Support Within the United States call 1 800 225 1480 outside the United States call 978 232 6000 or fax 978 232 6031 In Europe the Middle East and Africa contact your local authoriz...

Page 25: ...ctive Standard 89 336 EEC Electromagnetic Compatibility EMC Directive EMC essential requirements for measurement device EN 50081 1 1992 EN 55011 EN 50082 1 1997 EN 61000 4 2 3 These products have been...

Page 26: ...wing symbol Thermo Fisher Scientific has contracted with one or more recycling disposal companies in each EU Member State and this product should be disposed of or recycled through them Further inform...

Page 27: ...ific Environmental Instruments Water Analysis Instruments 166 Cummings Center Beverly MA 01915 USA Tel 978 232 6000 Toll Free 800 225 1480 Dom Fax 978 232 6015 Int l Fax 978 232 6031 www thermo com wa...

Reviews: