Thermo Scientific Russell RL060P pH Meter User Guide
0
pH Calibration using Pb standard buffers
1 . . Pour .a .known .pH .buffer .standard .solution .into .a .clean .container, .e .g . .pH .6 .97 . .
Power .on .the .meter, .and .it .automatically .enters .into .measurement .mode . .Select .
pH .mode .by .pressing .
. . .
2 . . Dip .both .pH .electrode .and .temperature .probe .into .pH .6 .97 .buffer .solution . .Swirl .
gently .and .wait .for .reading .to .stabilize .(about .30 .seconds .depending .on .your .
electrode .condition) . .
3 . . Press .
.to .enter .pH .calibration .mode . .A .“
CA
” .displays .momentarily .and .the .
display .shows .the .current .uncalibrated .reading .flashing .while .in .the .calibration .
mode .
4 . . To .abort .or .cancel .calibration .without .accepting .new .value, .press .
. .The .meter .
then .reverts .to .pH .measurement .mode .
5 . . To .proceed .with .the .calibration, .allow .reading .to .stabilize .first . .The .meter .
automatically .recognizes .either .pH .4 .10 .or .6 .97 .buffer . .Press .
.to .confirm .
calibration .and .the .LCD .displays .“
CO
” .momentarily . .The .meter .reverts .to .
measurement .mode . .
6 . . For .a .two .point .calibration, .repeat .steps .3 .through .5 .with .pH .4 .10 .buffer .for .better .
accuracy .