Thermo Scientific Russell RL060C Conductivity Meter User Guide
2
Select Cell Constant
This .meter .lets .you .select .a .cell .constant .of .K .= .1 .0, .10, .or .0 .1 . .
• . Use .a .cell .with .K .= .1 .0 .for .midrange .measurements
• . .Use .a .cell .with .K .= .10 .for .high .range .measurements .
(above .20 .mS) . . .
• . .Use .a .cell .with .K .= .0 .1 .for .low .range .measurements .
(below .20 .μS) . .
The .cell .included .with .your .meter .has .a .cell .constant . .
of .K .= .1 .0 .
1 . . .Enter .the .advanced .setup .as .described .in .the .
Advanced Setup Overview
section .
2 . . .Press .
.or .
.until .
CELC
.appears .on .the .LCD . .
Press .
. .
3 . . .Press .
.or .
.to .select .either .“1 .0”, .“0 .1” .or .
“10 .0” . .Ensure .the .cell .constant .selected .corresponds .
with .the .conductivity .probe .you .are .using . .
4 . . .Press .
.to .select . .The .meter .will .take .you .back .to .
the .menu, .
CELC
. .
5 . . .Press .
.or .
.to .move .to .the .next .menu .or .press .
.to .exit .to .measurement .mode . .