Thermo Scientific Russell RL060C Conductivity Meter User Guide
12
Selection of Automatic or Manual Calibration
This .meter .is .capable .of .performing .either .automatic .or .manual .calibration . .
In .the .automatic .calibration .mode, .the .meter .automatically .detects .and .verifies .the .
appropriate .known .calibration .standards .solutions .being .calibrated .before .accepting .
these .particular .calibration .standards .as .one .of .its .calibration .values .in .a .specific .
measurement .range . .This .automatic .calibration .mode .frees .you .from .cumbersome .
calibration .procedure . .
The .known .calibration .standards .used .for .automatic .calibration .are: .
Table 1
Conductivity .Calibration .Standards .for .Automatic .Calibrations
Meter
Normalization
Temperature
Calibration Standards (Range)
RL060C
25 .°C
1 .
84 .μS .(for .0 .– .200 .μS/cm)
2 . 1413 .μS .(for .0 .– .2000 .μS/cm)
3 . 12 .88 .mS .(for .0 .00 .– .20 .00 .mS/cm)
4 . 111 .8 .mS .(for .0 .0 .– .200 .0 .mS/cm) .
20 .°C
1 .
76 .μS .(for .0 .– .200 .μS/cm)
2 . 1278 .μS .(for .0 .– .2000 .μS/cm)
3 . 11 .67 .mS .(for .0 .00 .– .20 .00 .mS/cm)
4 . 102 .1 .mS .(for .0 .0 .– .200 .0 .mS/cm)
In .the .manual .calibration, .non-standard .calibration .values .can .be .used .for .calibration . .
You .can .manually .input .the .appropriate .values .as .your .desired .calibration .standards .in .
each .specific .range . .This .is .useful .when .you .have .a .customized .calibration .standard .
specifically .unique .for .your .application . .
To .select .automatic .or .manual .calibration .settings, .refer .to .the .
Automatic
Calibration
.section .for .more .information .