Thermo Scientific Russell RL060C Conductivity Meter User Guide
16
Temperature Calibration
The .conductivity .electrode .(Cat . .No . .014005) .has .a .built-in .temperature .sensor .for .ATC . .
The .temperature .sensor .is .factory .calibrated .to .the .meter . .Calibrate .your .sensor .only .if .
you .suspect .temperature .errors .that .may .have .occurred .over .a .long .period .of .time .or .if .
you .have .a .replacement .probe .
1 . . .Make .sure .that .the .phono .jack .(for .temperature .
measurement) .is .properly .connected .to .the .meter . .See .
Figure 7
. .
2 . . .Switch .on .the .meter .and .if .necessary, .press .
.
to .select .temperature .measurement .mode . .See .
the .
Change Conductivity to Temperature
Measurement Mode
section . .
3 . . .Press .
.to .start .the .temperature .calibration .process . .
4 . . .Dip .the .probe .into .a .solution .with .known .temperature .
(for .example, .a .temperature .bath) . .Allow .time .for .the .
temperature .to .stabilize . .
5 . . .Wait .for .the .value .to .stabilize .and .press .
.or .
.and .
adjust .the .value .to .the .solution .temperature .
6 . . .Press
. .The .[
CO
] .indicator .will .appear .for .1 .5 .seconds, .
and .the .reading .will .stop .flashing . .The .temperature .
calibration .is .successfully .performed . .The .meter .returns .
to .measurement .mode . .
Note:
To exit calibration without confirmation, press
.
again to go back to measurement mode.
Note:
You can offset the temperature reading up to ±5 °C
from the original (default) reading.
C
°
C
°
C
°
° C