Thermo Scientific Russell RL060C Conductivity Meter User Guide
1
6 . . .To .perform .the .next .point .calibration .in .the .multi-point .calibration, .repeat .steps .1-5 .
until .all .points .have .been .calibrated .
Important Notes:
1. Meter allows a tolerance range of ±40% of its calibration standard. An error
message “
Err 1
” will be displayed for 3 seconds if you attempt to calibrate with a
solution whose value is outside the tolerance range.
For instance: For 1413 μS conductivity calibration standard, 40% tolerance is from
848 μS to 1978 μS.
2. If the temperature (t °C) of the conductivity calibration solution is below 0 °C or
above 50 °C (0 °C < t °C > 50 °C), an error message “
Err 2
” will be displayed
when performing the auto calibration, and meter will return to measurement
mode.
3. All new calibration data will over-ride existing stored calibration data for each
measuring range calibrated.
4. It is important to use new conductivity calibration standards.
5. Low conductivity standard solution (less than 20 μS /cm) cannot be obtained
easily. Such low conductivity standard will be contaminated as soon as it is
exposed to air, therefore you must exercise caution during calibration in the first
measurement range (0.00 to 20.0 μS /cm).