Switching the Meter On
Press .
.to .power .up .your .meter . .
It .will .then .go .through .a .series .of .displays, .showing .the .various .setup .parameters . .
1 . . The .first .screen .shows .[
Con6
] . .
2 . . Second .screen .shows .[
C 1.0
] .which .is .the .conductivity .cell .constant, .k . .Default .
value .is .k=1 .0 . .You .can .select .different .cell .constants .of .0 .1, .1 .0 .or .10 .0 . .Refer .to .
the .
Advance Setup Functions
.section . .
3 . . Third .screen .shows .[
t 25.0 °C
], .which .is .the .normalization .temperature . .You .can .
set .the .normalization .temperature .to .either .25 .°C .or .20 .°C . .Refer .to .the .
Advance
Setup Functions
section . .Default .value .is .25 .°C . .
4 . . Fourth .screen .shows .[
t 2.1%
] .which .is .the .temperature .coefficient . .You .can .
customize .the .meter .with .a .temperature .coefficient .value .from .0 .0 .to .3 .0 .%/°C . .
Refer .to .the .
Advance Setup Functions
section . .Default .value .is .2 .1 .%/°C . .
5 . . All .LCD .segments .will .light .up .for .2 .seconds, .and .change .into .measurement .mode . .
6 . . You .are .now .ready .for .conductivity .measurement .