background image

14

 

|

 

ROSS Ultra, ROSS Ultra Triode and ROSS pH Electrodes

Thermo Scientific Orion

Buffer and Sample pH Changes

Buffer and sample pH values change with temperature because of 

their temperature dependent chemical equilibria. The pH electrode 

should be calibrated with buffers that have known pH values at different 

temperatures. Buffer values at different temperatures are given in  

Table 1.

Orion pH meters automatically calibrate with the correct pH buffer 

values based on the measured temperature. All pH meters are unable 

to correct sample pH values back to a reference temperature because 

every sample has a unique pH value versus temperature relationship. 

Therefore, calibration and measurements should be performed at the 

same temperature and pH values should be reported with temperature.

Reference Element Drift

Drift can occur when the internal reference elements inside the pH and 

reference portions of the electrode are reaching thermal equilibrium 

after a temperature change. Long-term drift or slow response can last 

until the sample and electrode are at the same temperature. 

ROSS electrodes use a unique internal reference to minimize the 

equilibration time.

Temperature Sensor Errors

When a pH and ATC probe are placed into a sample that varies 

significantly in temperature, the readings can drift for two reasons. First, 

the temperature response of the electrode and temperature probe may 

not be similar, which prolongs equilibration and drift. Second, a sample 

may not have a uniform temperature. Therefore, the pH electrode and 

temperature probe are responding to different environments. Stirring 

the sample will minimize this effect.

Using LogR technology, Orion 

V

ersa

 s

tar

 and Orion

 

D

ual

 s

tar

 

meters calculate the temperature by measuring the resistance value 

of the ROSS pH electrode’s sensing bulb. The pH and temperature 

response is identical and both measurements occur at the sensing 

bulb. Drift is minimized and errors due to environmental differences are 

eliminated.

Sure-Flow Electrode Leak Rate

The junction of the ROSS Sure-Flow electrodes has been designed 

to leak at a rate of approximately 1 mL per 24 hours. To keep the 

junction from clogging, the leak rate is slightly higher than conventional 

electrodes. During the average pH measurement time the loss of 

fill solution is minimal and the fill solution does not contaminate the 

sample.

ROSS Micro pH Electrode

The 8220BNWP ROSS micro pH electrode has been designed to 

measure extremely small samples. The measuring tip is 3 mm in 

diameter and 40 mm long. The minimum depth of immersion is 4.5 mm 

and the minimum sample size is 15 µL in 384 well plates. 

Summary of Contents for Orion ROSS

Page 1: ...ROSS Ultra ROSS Ultra Triode and ROSS pH Electrodes 263745 001 Revision B November 2014 Thermo Scientific Orion U ser Manual...

Page 2: ......

Page 3: ...ode Calibration 10 General Calibration Procedure 10 One Buffer Calibration 10 Two Buffer Calibration 11 pH Measurements 12 Electrode Storage 12 Electrode Maintenance 12 Electrode Cleaning Procedures 1...

Page 4: ...s are three to five times more precise than those obtained with conventional electrodes Over the lifetime of the electrode ROSS reference potential drifts less than an average 0 12mV per day which cor...

Page 5: ...ientific Orion Star A200 series Thermo Scientific Orion Star A100 series or Thermo Scientific Orion Dual Star meter with pH measurement capability such as the Versa Star benchtop meter with pH module...

Page 6: ...3 The electrode may arrive with the chamber partially filled with electrolyte solution Please fill the chamber with electrolyte solution before use Uncover the fill hole by removing the tape and add...

Page 7: ...an adequate flow rate the level of fill solution must always be above the reference junction and at least one inch above the sample level The fill hole should remain open whenever the electrode is in...

Page 8: ...on aqueous solutions and solutions that contain organic solvents A minimum of 20 water must be present in the sample for the best measurement results If there is a great deal of drift when using the R...

Page 9: ...a uniform rate Place a piece of insulating material such as polystyrene foam or cardboard between the magnetic stir plate and beaker to prevent measurement errors from the transfer of heat to the sam...

Page 10: ...the meter is set to automatically recognize the pH buffer at the measured temperature and an ATC probe is used you do not need to manually adjust the pH values of the buffers One Buffer Calibration 1...

Page 11: ...rode into the first buffer When the reading is stable set the meter to the pH value of the buffer at the measured temperature Refer to the meter user guide for a detailed procedure Table 1 provides pH...

Page 12: ...flush it with fresh ROSS fill solution and refill the chamber with fresh ROSS fill solution Electrode Cleaning Procedures General Cleaning We offer a variety of Orion pH cleaning solutions for your c...

Page 13: ...n Cat No 910001 for at least one hour Cleaning Solutions Cat No 900021 WA pH cleaning solution A for removing protein contaminants Cat No 900022 WA pH cleaning solution B for removing bacterial contam...

Page 14: ...C probe are placed into a sample that varies significantly in temperature the readings can drift for two reasons First the temperature response of the electrode and temperature probe may not be simila...

Page 15: ...m precision a three point temperature calibration is recommended Do not perform a one point temperature calibration if measured solutions will be below 20 C The following tables illustrate the expecte...

Page 16: ...s and at higher temperatures Because the pH membrane is composed of special low sodium error glass error due to sodium is negligible when measuring at pH values less than 12 When measuring at pH value...

Page 17: ...lectrode into fresh pH 4 buffer When the reading is stable record the mV value of the pH 4 buffer The mV value should be in the range of 150 to 210 mV 4 Calculate the absolute mV difference between th...

Page 18: ...Mercury Hg Cadmium Cd Sensor body with cable assembly X 0 0 Plastic hardware 0 0 0 Part Name Toxic or Hazardous Substances and Elements Hexavalent Chromium Cr VI Polybrominated biphenyls PBB Polybromi...

Page 19: ...combination pH electrode with epoxy body and semi micro tip waterproof BNC connector 8135BNUWP ROSS Ultra combination pH electrode with epoxy body and flat surface tip waterproof BNC connector 8156BN...

Page 20: ...ctrode with epoxy body waterproof BNC connector 8165DN ROSS Sure Flow combination pH electrode with epoxy body E DIN waterproof connector 8172BNWP ROSS Sure Flow combination pH electrode with glass bo...

Page 21: ...scientific com water for additional accessories buffers and buffer sizes Specifications pH Range 0 to 14 pH pH Precision 0 01 pH Temperature Range 0 to 100 C ROSS Ultra Triode Temperature Accuracy 1 C...

Page 22: ...8102BN 8104BN 8135BN 815600 8162SC 8165BNWP 8172BNWP 8302BNURCA 8107BNURCA 8157BNURCA 120 mm length 12 mm diameter 8103BNUWP 8103BN 8103SC 165 mm length 95 mm tip length 6 mm tip diameter 8115BNUWP 8...

Page 23: ......

Page 24: ...moscientific com water Netherlands Tel 31 020 4936270 info water uk thermo com China Tel 86 21 68654588 wai asia thermofisher com India Tel 91 22 4157 8800 wai asia thermofisher com Singapore Tel 65 6...

Reviews: