OdaLog Portable Gas Loggers
Operation | 17
Status
Figure 18. Status Screens
•
STATUS:
This screen scrolls through overview of
Instrument data.
•
Odalog G20:
This screen displays the variant (G20 or
G20 RTX) of Odalog G20 series.
•
SWVER:
This screen displays the Firmware version
number.
•
PEAK:
This screen displays gas level in ppm after Start
Log function.
•
CLOCK:
This screen displays (shown in 24-hour) time. If
clock is not set, this value may be incorrect on screen
and in instrument log. This must be set in the Odastat NX
Software.
•
TEMP:
This screen displays temperature (in Degrees
Fahrenheit / Celsius
–
depending on default units
selected in software).
•
BATTERY:
This screen displays battery level of the
Instrument.
•
MEMORY:
This screen displays the amount of local
memory available in GB.
•
TYPE OnCE:
This screen displays the type of logging
data.
•
SECOND:
This screen displays the logging interval i.e.
time between gas readings displayed as
“
S
ECONDS”
(set in the Odastat NX software).
•
CALGAS:
This screen displays the Calibration Gas Level
in ppm as set via the Odastat NX software package.
•
DUEEOL:
This screen displays
No of days for sensor
end of life.
•
DUECAL:
This screen displays
No of days for sensor
recalibration.
•
MODBUS:
This screen displays the status of the
communication. It is 0 when the instrument is not
connected to the software and 1 when the instrument is
connected through ModBus.
•
BPS:
This screen displays
the data transmission speed
from the instrument to the Odastat NX software
calculated in bits per second.
•
ALARMS
•
ALARMA:
Alarm Average
–
This screen displays the
high gas level alert from data averaged over a 12
hour rolling window.
•
ALARML:
Alarm Low
–
This screen displays the high
gas level alert, reaching the lower peak value set in
software.
•
ALARMH:
Alarm High
–
This screen displays the
high gas level alert, reaching the higher peak value
set in software.
•
SIGNAL:
This screen
displays the GSM signal strength of
the last successful transmission to the FTP server.
•
TRNSMT:
This screen
displays the transmission interval
set in the configuration (How often the data being pushed
to the server).
•
HUMD:
This screen
displays the internal humidity level
when the instrument is in the field.
ODALOG
G20
SWVER
1.5.6
PEAK
0.0ppm
CLOCK
12:50
TEMP
24.1
STATUS
°C
BATTERY
3.570
MEMORY
7GB
TYPE
OnCE
SECOND
60
CALGAS
33.0 ppm
DUEEOL
1670
DUECAL
252
MODBUS
0
bPS
ALARMA
10.0 ppm
ALARML
15.0 ppm
ALARMH
25.0 ppm
SIGNAL
34
TRNSMT
00:15
HUMD
48.5
INST
0.0ppm