12 | Operation
OdaLog Portable Gas Loggers
Status
Figure 9. Status Screens
•
Odalog L2:
This screen displays the variant
(G20 or G20 RTX) of Odalog G20 series.
•
SWVER:
his screen displays the Firmware version
number.
•
PEAK:
This screen displays gas level in ppm after Start
Log function.
•
CLOCK:
This screen displays (shown in 24-hour) time. If
clock is not set, this value may be incorrect on screen
and in instrument log
–
This must be set in the Odastat
NX Software.
•
TEMP:
This screen displays temperature (in Degrees
Fahrenheit / Celsius
–
depending on default units
selected in software).
•
BATTERY:
This screen displays battery level of the
Instrument.
•
MEMORY:
This screen displays the amount of local
memory available in GB.
•
TYPE OnCE:
This screen displays the type of logging
data.
•
SECOND:
This screen displays the logging interval i.e.
time between gas readings displayed as
“
S
ECONDS”
(set in the Odastat NX software).
•
CALGAS:
This screen displays the Calibration Gas Level
in ppm as set via the Odastat NX software package.
•
DUEEOL:
This screen displays
No of days for sensor
end of life.
•
DUECAL:
This screen displays
No of days for sensor
recalibration.
•
MODBUS:
This screen displays the status of the
communication. It is 0 when the instrument is not
connected to the software and 1 when the instrument is
connected through ModBus.
•
BPS:
This screen displays
the data transmission speed
from the instrument to the Odastat NX software
calculated in bits per second.
•
HUMD:
This screen
displays the internal humidity level
when the instrument is in the field.
SWVER
12.0.1
PEAK
0.0ppm
CLOCK
11:34
TEMP
16.2
BATTERY
3.503
°C
MEMORY
7GB
TYPE
OnCE
SECOND
900
CALGAS
33.0 ppm
DUEEOL
1670
DUECAL
239
MODBUS
0
bPS
HUMD
66.9
ODALOG
-L2-
INST
0.0ppm