OdaLog Portable Gas Loggers
General Specifications | 9
Note:
can be used.
Table 4. Battery Specifications
Model
Battery Type
Size
Quantity
Model
G20 RTX
Lithium
D
1
LSH20
G20
Lithium
D
1
LS33600
Table 5. Technical Specifications-Digital Sensor
Measurement
Value
Operating Principle
Electrochemical
Target Gas
H
2
S
Accuracy
±5% of measuring value
T50* Response Time
<15 seconds
T90* Response Time
<30 seconds
Output
ppm compensated for temperature
Sensor Warm Up Time
20 seconds
Overload (maximum concentration that can be
reached before damaging the sensor)
1000 ppm
Recovery Time
from 200 ppm to < 4 ppm
<180 seconds
Linearity*
Linear (±5% from linear, up to 200 ppm)
Measurement Interval
1 min
Resolution
0.1 ppm
Baseline Offset*
<±0.5 ppm H
2
S equivalent
Equilibrium Baseline Shift from -40°C to +60°C
<±3 ppm
Repeatability*
±5% of measured value
Orientation Sensitivity
None
Serial Communication
USB with Chip Select