Odalog G20 and G20 RTX
User Manual
XXXXXX
•
Revision A
November 2021
Page 1: ...Odalog G20 and G20 RTX User Manual XXXXXX Revision A November 2021...
Page 2: ...ientific It may not be copied or reproduced in any way without the express written consent of Thermo Fisher Scientific Thermo Fisher reserves the right to make changes to the product and this document...
Page 3: ...troduction 7 Odalog G20 Gas Logger 7 Odalog G20 RTX Gas Logger 7 General Specifications 8 Operation 10 Odalog G20 screens 10 Odalog G20 RTX screens 15 Modem and SMS 20 PC Software Odastat NX 21 Home S...
Page 4: ...s 39 Firmware Updater 40 Change Admin Password 40 Recommended Deployment 41 Signal Strength 41 Maintenance 42 Filter Replacement for G20 and G20 RTX 42 Sensor Calibration 43 Null Mode 43 Span Mode 43...
Page 5: ...st harsh environments Table 1 Applicable Models Model Range Part Number Resolution Communications Odalog G20 Gas Logger 0 200 ppm Odalog G20 H2S 200 0 1 ppm Modbus and Bluetooth 0 2000 ppm Odalog G20...
Page 6: ...talled and operated with a minimum distance of 20 cm between the radiator and all persons during normal operation Odalog G20 Gas Logger As the user of the product it is your responsibility to understa...
Page 7: ...d shall only be cleaned or wiped with a damp cloth when used in explosive atmospheres Only use the micro SD card that is provided or recommended by the manufacturer The enclosure shall only be disasse...
Page 8: ...disposal information is available from manufacture or contact distributor for information This instrument is not intrinsically safe and should not be used in potentially flammable or oxygen enriched a...
Page 9: ...ect To avoid build up of electrostatic charge the instrument shall be used or placed where it will be protected from inadvertent rubbing and shall only be cleaned or wiped with a damp cloth when used...
Page 10: ...d inspect the equipment within five days of delivery If you find any damage keep the packing materials and immediately report the damage to the carrier Do not return goods to the manufacturer without...
Page 11: ...ll the features from its respected predecessor adding new wireless data transmission and extended battery life to its already impressive repertoire With low power consumption coupled with a long life...
Page 12: ...able 4 Battery Life Run Time Greater than 12 months for life of Lithium battery Greater than 8 months depends on signal strength and server transmission interval Data Logging Capacity More than 10 mil...
Page 13: ...ring value T50 Response Time 15 seconds T90 Response Time 30 seconds Output ppm compensated for temperature Sensor Warm Up Time 20 seconds Overload maximum concentration that can be reached before dam...
Page 14: ...ttery percentage as shown in Figure 5 Displaying battery percentage is common for all screens 4 Communications Symbol This symbol indicates if the instrument is successfully connected and transferring...
Page 15: ...t until the display counts down from OFF2 to OFF Then release button When the button is correctly released to turn the instrument off the OFF title will shift 1 character to the right to confirm that...
Page 16: ...ng data SECOND This screen displays the logging interval i e time between gas readings displayed as SECONDS set in the Odastat NX software CALGAS This screen displays the Calibration Gas Level in ppm...
Page 17: ...communication and power supply interfacing between computers peripherals and other computers In Odalog G20 it transfers data between the instrument and the Odastat NX software The device is the trans...
Page 18: ...efly on the screen Span Figure 13 Span Screen Hold the Magnet next to the Menu until display progresses to SPAN and then release button SPAN IN Prog will appear to indicate the instrument is in Span m...
Page 19: ...enables logging START LOG This screen enables logging the data This is confirmed by LOGGING symbol flashing on screen USB This screen enables USB Universal Asynchronous Receiver Transmitter link for c...
Page 20: ...T Last is displayed and then release MODEM In It This screen displays the Instrument powering up the modem Instrument will return to the INST screen The LOGGING symbol on screen should now be flashing...
Page 21: ...end of life DUECAL This screen displays No of days for sensor recalibration MODBUS This screen displays the status of the communication It is 0 when the instrument is not connected to the software and...
Page 22: ...wer supply interfacing between computers peripherals and other computers In Odalog G20 RTX it transfers data between the instrument and the Odastat NX software The device is the transmission line and...
Page 23: ...n the screen Span Figure 22 Span Screen Hold the Magnet next to the Menu until display progresses to SPAN and then release button SPAN IN Prog will appear to indicate the instrument is in Span mode Us...
Page 24: ...pecifications Explicit Connection on Port 21 Should support TLS 1 2 AES 265SHA Table 7 Test SMS 12345 SITE ID BATT 100 EOL 1664 CALDUE 95 AVAMEM 14GB Alarm Average SERIAL 12345 SITE ID High Gas Alert...
Page 25: ...s Loggers PC Software Odastat NX 21 PC Software Odastat NX Home Screen The main functions of Odastat NX software are easily accessible on the Start screen See Figure 23 for reference Figure 23 Odastat...
Page 26: ...op down and Click on Connect to connect to the device User can configure the device and download the data using an USB or Bluetooth Figure 24 Select Connection Screen Select the relevant Communication...
Page 27: ...Select the Bluetooth Communication screen from the Select Connection tab at the top right end of the screen Click Connect button Pair the instrument if you are connecting for the first time alternati...
Page 28: ...tion Device Paired User get a Successfully Connected and Paired message as shown in Figure 29 after the device is successfully connected Figure 29 Bluetooth Communication Successfully Connection Once...
Page 29: ...le Gas Loggers PC Software Odastat NX 25 Figure 30 Bluetooth Communication Login Screen After Login user can set the date and time from the Set Date and Time Figure 31 Bluetooth Communication Date and...
Page 30: ...Login Screen Enter your Username and Password to Login into your Odalog account Figure 32 Odalog Login Screen If the Username and password are correct then a Successfully logged in message appears on...
Page 31: ...me screen from the Odalog main menu Select the time from Set Date and Time screen from the Odalog main menu Figure 34 Date and Time Screen Note User needs to set the unit time every time they replace...
Page 32: ...x Alarm Settings These controls allow the user of the instrument to set low average and high gas concentration threshold values When these values are exceeded the instrument will notify the user via e...
Page 33: ...ta via Modbus to a Modbus Master the following parameters are to be filled Instrument Parameters ModBus Register Start Address Device ID Baud Rate Odalog G20 supports following baud rates i 9600 ii 19...
Page 34: ...em Enter the relevant value Modem Wakeup Interval This is time interval for data Transmission Enter the relevant value Mobile Number Enter the mobile number to get the alerts FTP Enter the relevant Se...
Page 35: ...find Device Details The screen displays Model Firmware version Sensor Selection and the Serial Number of Odalog Figure 38 Instrument Parameters Manufacturing Modify User Details Click the Modify User...
Page 36: ...Parameter File Download Log Screen After successfully connection select Download Log from the Odalog main menu The program will immediately start to download the currently connected Odalog s logged d...
Page 37: ...nt has finished downloading user will receive the message that the data is saved Figure 42 After Downloading data Clear Log This sends a signal to the Odalog instrument to clear the entire log Operati...
Page 38: ...re 44 After clicking the button the Open File Selector will be displayed This open dialog is unique to Odastat NX and allows the user to see information about the logged files without opening them Fig...
Page 39: ...his file as a Graph Figure 46 Session File Selection Chart Screen When a previously stored log file is opened it can be displayed in a Graph Format Selected sessions of data collected from different d...
Page 40: ...s Figure 47 and Figure 48 show sessions of data collected from different time periods Figure 48 Sensor Data Chart September The parameters in the graph can also be displayed in a Seperate Mode as show...
Page 41: ...instrument is to be established Odalog G20 RTX communicates with the software through internet using File Transfer Protocol To establish communication with the instrument G20 or G20 RTX use the Conne...
Page 42: ...ow the customer to select a folder containing data transmitted by each Odalog G20 RTX Select the respective Odalog folder See Figure 52 Figure 52 Odalog Folder Selection Choose the appropriate folder...
Page 43: ...54 Online Parameters File Preferences The File Preferences are used to change default settings of the file save destination directories that OdaStat will use to save files as shown in Figure 55 Click...
Page 44: ...OdaLog Portable Gas Loggers Firmware Updater The admin or the user can update the firmware in the Firmware Updater screen Figure 56 Firmware Updater Change Admin Password Figure 57 Production Tool Scr...
Page 45: ...nsor for the sensor to respond It s common for the Odalog to be placed into areas covered by thick concrete and or metal covers It is important however that the signal can still transmit when deployed...
Page 46: ...the part is listed as a replacement part Only use Thermo Fisher International replacement parts The Odalog should be switched off when not in use and stored in a clean and dry environment as usually...
Page 47: ...t plastic tubing over an inlet on the calibration adaptor Figure 59 Open regulator and allow gas to pass over the sensor at the flow rate and for the flow time as detailed in Table 3 After waiting for...
Page 48: ...de 103 Electrolyte dry out SENSOR bAd2 Error code 113 Waterlogged SENSOR bAd3 Error code 109 Span Calibration is due SENSOR bAd4 Error code 104 End of life SENSOR bAd5 Error code 001 Diagnostic electr...
Page 49: ...om In New Zealand For customer service call 0800 933 966 To fax an order use 0800 329 246 To email an order ordersnz thermofisher com Find out more at www thermofisher com 2021 Thermo Fisher Scientifi...