- 5 -
Section II General Information
Description
The DIMAX Liquid to Liquid Heat Exchanger is designed to remove heat
loads from liquid-cooled instruments. The unit uses facility cooling or tap
water as the secondary cooling medium to remove heat from the primaryfluid
in the closed recirculation loop.
The unit consists of a plate heat exchanger, recirculation pump, stainless
steel reservoir, built-in replaceable deionizing cartridge, and a microprocessor
controller. The controller monitors temperature, controls resistivity and
displays all fluid flows.
The unit has 11 return lines. The flow rate in 8 of these lines as well as the
facility flow line is monitored by flow transducers. The unit also has fluid level
sensors designed to protect the unit's pump and your application. A resistivity
sensor monitors resistivity and sends a signal to the controller when replace-
ment is necessary.