- 14 -
Configuration Loop
The configuration loop allows you to configure frequency, enable the buzzer
and verify software version number.
To enter this loop you must be in the operators loop and displaying the
temperature. Depress and hold the ENTER key. While holding the ENTER
key enter the key sequence NO-YES-NO.
Scroll through the CONFIGURE menu using the NEXT or LAST key, press
ENTER to go to the desired sub-menu. Scroll through the sub-menu options
using the NEXT or LAST key, toggle the options using the YES or NO key.
Leave the sub-menu using the NEXT or LAST key. The display will indicate
the save prompt, press YES or NO.
Figure 5 Configuration Loop