- 21 -
DeviceNet
Interface
Resistivity Scaling
Unsigned binary is used to represent the 0V to +10V range. 12 bits of resolu-
tion is available at the DeviceNet interface.
0 to +10V range
AI Digital Value
Resistivity input to controller
(at the DeviceNet Interface)
0
0 Mohm
4095
20 Mohm
Flow Scaling
Unsigned binary is used to represent the 0V to +10V range. 12 bits of resolu-
tion is available at the DeviceNet interface.
Facility Flow
0 to +10V range
AI Digital Value
Flow input to controller
(at the DeviceNet Interface)
0
0 gpm
4095
100 gpm
Process Flow
0 to +10V range
AI Digital Value
Flow input to controller
(at the DeviceNet Interface)
0
0 gpm
4095
20 gpm
Temperature Scaling
Unsigned binary is used to represent the -5V to +5V range. 12 bits of resolu-
tion is available at the DeviceNet interface.
-5V to +5V range
AI Digital Value
Temperature input to controller
(at the DeviceNet Interface)
0
0°C
2048
+50°C
4095
+100°C