Preventative maintenance procedures
32
Thermo Scientific™ Multidrop™ Pico 8 Digital Dispense Operating Manual
Thermo Fisher Scientific
A cassette and plate are loaded in the first step. Any type of cassette can
be used for alignment. Place a clear lid or plate seal on the plate.
After the cassette and plate are loaded, click Next and they will be
validated. If necessary, you will be prompted again for cassette and/or
plate load.
DMSO or an aqueous fluid with surfactant can be used for dispensing.
Colorant (not more than 10% food coloring in DMSO) may be added
to the fluid to make it easier to verify alignment.
After loading the fluid and clicking Next, fluid will be dispensed to the
centers of the 4 corner wells.
After dispensing, the plate will be moved to a position where it can be
examined or removed for closer examination.
Load and
validate
Fluid and
Dispense
Adjust