Cleaning and decontamination
9.7
Cleaning the floorpan
Clean the floorpan using a solution of tepid water and commer-
cially available dishwashing agent.
1. Remove the workplates from the sample chamber.
2. Remove dirt residues and deposits thoroughly.
3. Wipe the floorpan clean using a clean cloth and plenty of
clear water.
4. Remove the cleaning liquid from the floorpan and wipe the
floorpan surfaces thoroughly clean.
NOTE – Material residues!
After cleaning, make sure that all cleaning
materials have been removed completely from
the floorpan.
5. Reinstall the workplates.
9.8
Cleaning the protective grid
Fig. 16:
The multi-segment protective grid is retained in the
installed position between support surface and airduct wall by
its own tension.
1. To remove a grid segment [2], push the retaining tab [1]
down until the grid segment can be removed from below
the counterholder [3] of the wall.
2. To install the grid segment [2], first place it behind the fixed
points [4], then push the retaining tab [1] down and against
the counterholder [3] of the wall so that the counterholder
engages in the tab.
NOTE – Inlet air protection!
Do not operate the device without inlet air
protection. Prior to any start-up of the device,
make sure that the inlet air protection is in-
stalled!
Fig. 16
Protective grid segment